X-ray Tube

HOME > PRODUCTS > TUBES

  OX/110-0514  
DESCRIPTION AND TECHNICAL DATA
Voltage
nominal
120 kV
maximum for test
135 kV
Inverse voltage
nominal
120 kV
maximum for test
135 kV
Focal spot (IEC60336:2005)
0.5 mm - 1.4 mm
Filament characteristics
small 2.5 ÷ 4.0 V 3.0 ÷ 4.0 A
large 6.5 ÷ 12.5 V 3.0 ÷ 4.5 A
Nominal anode input power at 0.1 s (DC)
small 1200 W
large 4500 W
Anode material
tungsten
Target angle
15 °
Anode heat storage capacity
33000* J
Maximum anode cooling rate
600 W
Maximum anode cooling rate
280* W
Inherent filtration
0.5 mm Al
Maximum diameter
50 mm
Overall length
162* mm

* with std shank
DIMENSIONS AND CONNECTIONS